Recirculating chiller: Mini and ECO Mini from ATC – cooling capacity: 500 W

They are low-noise small active air/water continuous chillers for low cooling capacities and accurate temperature control (+/- 0.1 °C).

ATC is the UK’s leading manufacturer of recirculating chillers and temperature control products. From standard products to fully customized systems, ATC can offer you the best cooling solution and temperature control for your problems.

What distinguishes the devices:

  • Small – like a desktop PC size
  • Quiet – variable fan speed and demand-driven compressor
  • Ideal for laboratory applications. Due to the programmable temperature curve (timetabel and setpoints), the mini cooler shows its strengths here.
  • Very suitable for industrial applications. Due to the programmable RS-485 interface, up to 247 mini coolers can be individually controlled together via one system.
  • Eco Mini is the more cost-effective version with limited control.

Use for cooling / temperature control of:

  • Bioreactors
  • Elektronenmikroskope
  • Laser
  • Vacuum pump
  • Rotary evaporators
  • Elektrophoresis chambers
  • Water baths

Relevant performance specifications

MiniECO Mini
Cooling capacity1500 Watt500 Watt
Temperature range
(standard)
+ 4 °C to + 35°C+ 4 °C to + 35°C
Temperature range
(optional)
– 10 °C to + 35°C– 10 °C to + 65°C
Temperatur
stability at constant load
+/- 0.1 °C+/- 0.1 °C
Principlecompression chiller

active air / water chiller
compression chiller

active air / water chiller
pump options and flow ratesP7 – 7 l/minP7 – 7 l/min
System volume0.64 l0.64 l
CompatibilityHexid, propylene-glycol, destilled waterHexid, propylene-glycol, destilled water
Operating pressure range0.4 to 1.6 bar0.4 to 1.6 bar

1 Performance at 20°C ambient temperature at the chiller and 20°C setpoint for application to be cooled

Other relevant specifications

MiniECO Mini
Weight20 kg20 kg
Size (L x W x H) 435 x 180 x 380
mm
435 x 180 x 380
mm
Noise level64 dB (A) 1m35 – 64 dB (A) 1 m
F-gas amount2 /
type, GWP1 and
CO2 equivalent
0.36 kg R134a
(GWP 1430)
0.52 t CO2
0.36 kg R134a
(GWP 1430)
0.52 t CO2
Operating voltage and
maximum power
requirement / for 50 Hz
230 V – 1 A230 V – 1 A
Environmental
temperature
+ 5 °C to + 40 °C+ 5 °C to + 40 °C
Min. inside diameter
of pipeline
8 mm8 mm

1GWP: Global warming potential – the global warming potential that this refrigerant has compared to CO2 over a period of 100 years